TUBE REPLACEMENT CHECKLIST
Hologic Dimensions • Tube Replacement Checklist
Service Information
This is a field checklist based on tube replacement notes. Follow OEM service documentation, radiation safety procedures, site policy, and local regulatory requirements.
Correct Calibration Order After Tube Change
1. Before Removing or Moving the Old Tube
GoalGet the new tube into the same physical position as the old tube.
Put the new tube in the middle as best you can. Center it left-to-right and front-to-back.
2. New Tube Installation Details
The locking collar screw must not be directly on top of the tube. If it is on top, the top cover may not fit flush when reinstalled.
3. Tube Warmup
Use the CalTool version for that system. Do not use CalTool 2.2 unless that is the correct version for the unit.
Use Step 2 to take manual shots in the 25–28 kV range.
Do not spend time on Step 3 for this process.
4. Large Filament Calibration – 35 kV
Large Filament Calibration reloads defaults to the generator node, which gives a better starting point for a new tube.
For this workflow, perform the 35 kV large filament calibration only.
Toward the end, CalTool may throw an error saying it reached the maximum number of exposures.
When the calibration completes, hit the Apply button.
5. Generator Calibrations
Only start these after Large Filament 35 kV is completed and applied.
6. Small Filament Calibration – 30 kV
7. HVL / TO Calibration
The first exposure it asks you to make is for the small focal spot. That reading goes into the box on the right, which is the mR reading.
1. The form. 2. CalTool. 3. The Piranha.
The Al filter tomo shots are done on the next step after the Ag sequence.
For tomo shots, if the mR goes over 1000, the meter may switch and display the value as R. You still need to enter it into the form as mR. Example: 1.3 R = 1300 mR.
HVL / TO Quick Reference
| Step | What To Do | Important Reminder |
|---|---|---|
| First Exposure | Small focal spot exposure | Enter into the right-side mR box |
| Rh | Do the 22 kV Rh shot, then continue Rh sequence | Complete Rh before switching filters |
| Ag | Switch to Ag and perform Ag shots | Change form, CalTool, and Piranha |
| Al / Tomo | Perform tomo shots on the next step | If reading displays R, convert to mR |
| Cu | Only perform if licensed | System may not have a Cu license |
8. Licensed Filters and Output Form
The system may not have a Cu license. Do only the filters that are licensed for that system.
Text a picture of the output form for review.
9. Tube Head Check / Cooldown
You may want to let it cool down a bit before continuing.
10. AEC Calibration Start
Start with Dark Offsets, then Goaltab Gen, then Goaltab Scale.
Additional AEC steps may be required depending on the next instructions and system condition.